Optimizer Lab Scale Filters

Small volume lab scale filters
 
  • Advanced filtration media: optimizes filtration and particle retention capabilities
  • Easy changeout design: allows for quick particle analysis on the membrane
  • R&D laboratory use with scale-up capability: lower cost membrane evaluation
 
Optimizer C47 Housing and Membrane Disks
Membranes include:
Ultra high molecular weight polyethylene (UPE) (hydrophobic)

PCM (hydrophilic)

Protego® polyethylene ion exchange media

Polypropylene media
Membrane area:
12 cm2
Housing materials:
Head, Bowl, Locking Ring: High density polyethylene

O-rings: EPDM or Perfluoroelastomer
Bowl size:
2-136
Head size:
2-223
Hold-up volume:
8.5 mL
Connections:
In/Out/Vent: 1/4” NPT
Maximum operating conditions:
Maximum operating pressure: Perfluoroelastomer O-ring: 0.21 MPa (2.1 bar, 30 PSI) at 25°C; EPDM O-ring: 0.28 MPa (2.8 bar, 40 PSI) at 25°C

Maximum operating temperature: 40°C
Optimizer V47 Filters
Materials:
Membrane: Patented ultra high molecular weight polyethylene (UPE)

Housing: Polypropylene

Supports: Polypropylene non-woven material
Hold-up volume:
1.5 mL
Membrane area:
13.4 cm2
Maximum operating conditions:
Maximum operating pressure: 0.40 MPa (4.0 bar, 58 PSI) at 20°C