• our science our science Our Science
    • By Industry

      • Life Sciences

      • Microelectronics

        • Semiconductor

        • Data Storage

        • Flat Panel Display

        • LED

        • Solar

      • Industrial

        • Aerospace

        • Chemical Manufacturing

        • Chromatography and Analytical

        • Coatings, Inks, and Adhesives

        • EDM

        • General Industrial

        • Glass Forming

        • Polymer Manufacturing

        • Water Treatment

    • By Role

      • Logic Manufacturers

      • Memory Manufacturers

      • Semiconductor Equipment Manufacturers

      • Chemical Suppliers

      • Bioprocessors

    • By Solution Area

      • Contamination Control

      • Fluid Management

      • Life Sciences

      • Specialty Materials

      • Substrate Handling

    • By Service

      • Analytical Services

      • Field Support Services

      • FOUP Services

      • Refurbishment Services

      • Regeneration Services

      • Technology Centers

        • China Technology Center

        • Life Sciences Technology Center

  • product catalog product catalog Product Catalog
    • Bare Die/CSP Trays

      • Tray Wizard

      • H20 Series Accessories

      • H44 Series Accessories

    • Chemistries

      • Chemical Delivery Systems

        • Solid Chemical Delivery Systems

        • Liquid Chemical Delivery Systems

      • Specialty Chemicals

        • Catalysts

        • Coating, Ink, and Adhesive Ingredients

          • Resin Monomers and Components

          • Surface Modifiers

        • Post-CMP Cleaning Solutions

          • Semiconductor Cleaning Solutions

            • PlanarClean® Cleaning Solutions

            • ESC Cleaning Solutions

          • Wide Band Gap Semiconductor Cleaning Solutions

          • HDD Media Substrate Cleaning Solutions

        • Post Etch Cleaning Solutions

          • TitanKlean® Cleaning Solutions

          • ST Cleaning Solutions

          • NOE Etchant Cleaning Solutions

        • Precursors

          • Upstream ALD/CVD Precursors

          • Advanced Deposition Materials (ADM) ALD/CVD Precursors

        • Electronic Chemicals

        • Slurries

        • Silanes

        • Organophosphorus Compounds

        • Specialty Acrylates and Methacrylates

        • Siloxanes

        • Specialty Additives

        • Ligands

        • Wafer Reclaim

    • Fluid Management Systems

      • Fittings Selection Tool

      • Liquid Packaging

        • Chemical and Bioprocessing Containers

          • FluoroPure® HDPE Blowmolded Drums

            • FluoroPure® Advantage Trilayer HDPE Blowmolded Drums

            • FluoroPure® Trilayer HDPE Blowmolded Drums

          • FluoroPure® Composite Drums

          • FluoroPure® Custom Products

        • Single-Use Assemblies

          • 2D and 3D Bags

          • Mixing Systems

          • Motion Bioreactor Bags

          • Microcarrier and Cell Separation Systems

          • Secondary Containment

        • NOWPak® Liner Based Systems

        • Sentry® Quick Connect Systems

        • FluoroPure® Intermediate Bulk Containers

        • FluoroPure® Port Options, Tools and Accessories

        • Tube Sets and Manifolds

        • Custom Dip Tubes, Tanks, and Containers

        • FluoroPure® Pressure Vessels

        • Process Tanks

      • Valves Selection Tool

      • Fluid Handling

        • Fittings

          • PrimeLock® Tube Fittings

          • PrimeLock® ESD Tube Fittings

          • PrimeLock® Accessories

          • Flaretek® Tube Fittings

          • Flaretek® Accessories

          • Cynergy® Fittings

          • Cynergy® Accessories

          • EP PFA Flare Fittings

          • EP PFA Flare Fitting Accessories

          • PureBond® Welded Tube Fittings

          • Quikgrip® Tube Fittings

          • Quikgrip® PFA Nuts

          • Integral Ferrule Tube Fittings

          • Integral Ferrule Accessories

          • Barb PFA Tube Fittings

          • Dual Containment Fittings

          • PureBond® Welded Pipe Fittings

          • PureBond® Accessories

          • NPT Pipe Fittings

          • NPT Accessories

        • Valves

          • CR and CH Series Valves

            • CR4 Series Valves

            • CR8 Series Valves

            • CH8 Series High-Temperature Valves

          • Integra® Valves

          • Cynergy® Valves

          • EP PFA Valves

          • Manifolds and Assemblies

          • Plug Valves

          • Needle/Metering Valves

          • Stopcock Valves

          • Check Valves

          • Valve Accessories

        • Fluid Handling Accessories

          • Spray Products

          • Aspirators

          • Fasteners

        • Valves Selection Tool

        • Fittings Selection Tool

        • Tubing and Pipe

        • Custom Products

      • Process Monitoring

        • Process Control

          • Integrated Flow Controllers

          • NT™ Proportional Control Valves

          • Photochemical Dispense Pumps

        • Process Measurement

          • Concentration Monitors

          • Electronic Flowmeters

          • NT™ Pressure Transducers

        • Particle Characterization

    • Gas Filtration and Purification

      • Gas Diffusers

        • Chambergard™ Gas Diffusers

      • Gas Filters

        • Other Gas Filters

        • In-Line PTFE Gas Filters

        • In-Line Stainless Steel Gas Filters

        • In-Line Nickel Gas Filters

        • Surface Mount Gas Filters

      • AMC Filters

        • Stepper and Scanner Filters

        • Process Tool Air Filters

        • Facility Filtration

        • Filter Housings and Stand-alone Filter Cabinets

      • Gas Purifiers

    • Hard Disk Component Handling

      • Storage Boxes and Individual Disk Packages

      • Read-Write Trays and Carriers

      • Disk Shippers

      • Disk Process Carriers

    • Liquid Filtration and Purification

      • Liquid Filters

      • Liquid Purifiers

      • Liquid Filter Housings

    • Mask and Reticle Handling

      • Reticle Pods and Purge Cabinets

      • Mask Carriers and Shipping Boxes

      • Mask Packages

    • Other Substrate Handling

      • Miscellaneous Device Processing

      • Solar Cell Solutions

    • Specialty Materials

      • Premium Graphite

        • Industrial Graphite Grades

        • Semiconductor Graphite Grades

        • Life Sciences Graphite Grades

        • Premium Graphite Components

        • Glass Forming Graphite Grades

        • GLASSMATE® Components

        • Graphite Processing Options

      • Coatings

        • Chucks

        • Plasma Enhanced Chemical Vapor Deposition (PECVD) Coatings

        • Caerus™ Coating Technology

        • Pegasus™ Coating Technology

      • Premium Silicon Carbide

        • SUPERSiC<sup>®</sup> Silicon Carbide Components

        • SUPERSiC<sup>®</sup> Silicon Carbide

      • Advanced Cleaning Materials

    • Specialty Gases

      • Specialty Gas Mixtures

      • Gas Delivery Cylinder Systems

      • Gas Delivery Cabinet Systems

      • Gases

    • Wafer Handling

      • Wafer Processing

        • CMP Cleaning Brushes

        • CMP Pads

        • 300 mm Front Opening Unified Pods (FOUPs)

        • 200 mm Wafer Processing

        • 200 mm Accessories

        • 150 mm Wafer Processing

        • 150 mm and Smaller Wafer Carrier Accessories

        • 125 mm Wafer Processing

        • 100 mm Wafer Processing

        • 76.2 mm 3" Wafer Processing

        • 2.5" and Smaller Wafer Processing

        • Labware

        • Chucks

      • Wafer Shipping

        • 150 mm Wafer Shippers

        • 125 mm and Smaller Wafer Shippers

        • 300 mm Wafer Shippers

        • 200 mm Wafer Shippers

      • Finished Wafer Shipping

        • SmartStack® Contactless Horizontal Wafer Shipper

        • SmartStack® Horizontal Wafer Shippers

        • Film Frame Rings and Shippers

  • resources resources Resources
    • Reference Materials

    • Technical Information

      • Calculators and Tools

      • Chemical Compatibility

      • Chemlock® Filter Housing Technical Information

      • Component Technical Information

      • Entegris Test Standards and Protocols

      • FAQ

      • Formulas

      • Glossary

      • Material Properties and Safety

      • Particle Characterization Applications

    • Industry Insights

      • 3D NAND Solutions

      • Achieving Hydrogen Purity Standards for Fuel Cells

      • Advanced Coatings

      • Advanced Deposition Materials Innovation

      • Automotive Reliability

      • Cars and Chips: The Acceleration of Electronic Systems

      • Chasing the Perfect Pattern

      • Chemical Mechanical Planarization (CMP)

      • Clean Chemical Delivery

      • Clean Gas Delivery

      • CMP Process Monitoring

      • Cold Chain Bioprocessing

      • Dual Containment Tubing

      • Electrostatic Discharge (ESD) Prevention

      • Enabling Advanced Lithography

      • End-to-End Gas Filtration Solutions

      • Food and Beverage Contamination Control

      • Front-end to Back-end Solutions

      • Glass in Front; Glass in Back

      • High-Density Cell Banking

      • High Purity Chemical Manufacturing

      • Holistic Approach to Enabling Device Performance, Yield, and Reliability

      • Internet of Things

      • Low-Temperature Bioprocessing

      • More Than Just a Chip Tray

      • Operational Excellence

      • Pandemic Response

      • Particle Characterization

      • Pervasive Defectivity in Semiconductor Manufacturing

      • Protecting Your Data Center Environment from Gas-Phase Contamination

      • The Rare Air of the Semiconductor Fab

      • Silicon Carbide CMP Solutions

      • The Silicon Precursor Toolbox for Low-Temperature Deposition

      • Solid Precursors and Delivery Systems

      • Solutions in Advanced Logic

      • Solving Advanced Technology Challenges with Innovative Materials

      • Storage Class Memory

      • Targeted Contaminant Removal

      • Specialty Gases and Delivery Systems for Ion Implantation

      • Accelerating Speed-to-Yield with High-Purity Materials Solutions

    • Ensights (Blog)

    • Scientific Report

  • about us about us About Us
    • Corporate Overview

    • Corporate Social Responsibility

    • Careers

    • Events

    • Investor Relations

    • Legal/Trademark Notices

    • Locations

    • News Releases

    • Media Coverage

    • Entegris Blog

    • Manufacturing

Two Stage Dilusion


To deal with extremely concentrated colloidal suspensions, Particle Sizing Systems developed a patented two-stage Autodilution scheme. Besides being able to perform dilutions up to 200,000x within minutes, the two-stage dilution scheme allows for automatic sampling through the use of a sample-loop capture system. This extends the capabilities of the Single Particle Optical Sizing (SPOS) technique to online analysis and the addition of an autosampler.

Certain industrial colloidal systems like paints and inks are extremely concentrated and require a lot of dilution in order to employ SPOS. The figure displays a schematic of the Two-Stage Autodilution scheme as it would be used in the AccuSizer 780 APS.

The two-stage Autodilution scheme uses a two position fill-inject valve which contains a loop of known volume, V1. In the fill position, sample is allowed to flow through the loop and out the valve thereby filling the loop with concentrated sample. 

The sample could be under pressure from a reactor or process stream or a pump could pull sample through. After the loop is full, the valve is put into the inject position allowing pump 1 to flow diluent into the loop at flow rate F1, carrying the sample into a dilution chamber of known volume, V2. It is important to note that pump 1 can generate variable flow rates. This accomplishes the first stage of dilution. The dilution factor of this stage is V2/V1. Typical values for V1 and V2 are 0.5 ml and 30 ml, leading to a first stage dilution factor of 60X.

Simultaneous to this, pump 2 starts to flow clean diluent into the static mixer at flow rate F2. F2 is much larger than F1. Even after the sample is injected into the dilution chamber, pump 1 continues to operate, flowing in clean diluent into the dilution chamber. Since the dilution chamber is closed, liquid exits into the static mixer at the same flow rate as pump 1. There it mixes with the flow from pump 2. This represents the second stage of dilution. The dilution factor can be expressed as:

2nd stage dilution factor = 1+ F2/F1

As can be seen from the above equation, the second stage dilution can be controlled by changing the flow rate of pump 1. In this two-stage scheme, instead of diluting continuously like in the exponential Autodilution arrangement, the flow rate of pump 1 can be increased or decreased to obtain a steady-state count rate, as monitored by the light obscuration sensor and set by the user. That is why this fluidics arrangement is ideal for online analysis. As might be expected by a process stream, the concentration of particles will change with time. The two-stage dilution scheme can adjust the overall dilution factor as required from measurement to measurement as the sample concentration changes. This two-stage mixed-flow method of dilution not only provides a large range of dilution factors that can be adjusted based on the concentration of sample, but once a set point concentration is achieved, sample can be delivered to the SPOS sensor at a nearly constant concentration for relatively long periods of time. This permits a large number of sample particles to be counted, resulting in enhanced statistical accuracy of the resulting particle size distribution.

A count rate profile for a two-stage Autodilution measurement is shown:

For this measurement, the set point concentration was 7,000 particles/ml. The first F1 the fluidics settled on produced a particle concentration of 17,000 particles/ml. The search algorithm lowered the F1 flow rate, thus increasing the 2nd stage dilution factor, which produced a particle concentration of 7,060 particles/ml, within range of the set point. This was the F1 used for the actual data collection. As can be seen, the particle concentration remained constant throughout the entire measurement.

Just like in the Exponential Autodilution scheme, the concept of Dilution Factor can be utilized. As before, Dilution Factor or DF is defined at the ratio of particles introduced into the system over the number of particles counted in the measurement. Phrasing it in terms of the two-stage dilution scheme, DF would be the number of particles in the loop over the number of particles counted in a measurement. Since the volume of the loop and the number of particles counted are known and the DF is measured, the particle concentration can be expressed as:

Co= (Particles Counted x DF)/V1

Keeping in mind the expressions for the first stage dilution factor and the second stage dilution factor, the following equation for DF can be developed:

DF = V2/(F1x T)

Where T is the data collection time

This is a simplified expression as it assumes that the particle concentration in the dilution chamber will not change with time. However, this is not always true. As pump 1 is continuously flowing clean diluent into the dilution chamber, the particle concentration in the dilution chamber falls exponentially (see Autodilution ). In the case of the extremely concentrated colloidal dispersions that the two-stage Autodilution system would be employed on, this exponential decrease is extremely slow and the above expression for DF holds. In the case where the concentration will fall off significantly during the data collection time T, an exponential correction factor needs to be added into the expression for DF.

An example of a two-stage Autodilution measurement can be seen below:

The particle size distribution is from a commercial dispersion of 1 micron polystyrene latex spheres. These particles are very narrowly dispersed. They were introduced into the system as a 1:1000 predilution which is easy to make on the laboratory bench. The measured particle concentration was 1.83 x 107 particles/ml well within the expected particle concentration of 1.9 +/- 0.2 x 107 particles/ml. The fluidics were setup with a 1 ml sample loop volume (V1= 1 ml). The software chose a flow rate for pump 1 of 0.65 ml/min (F1= 0.65 ml/min). With an F2= 60 ml/min, the second stage dilution factor will be 93.4. The number of particles counted in a 120 second collection time was 807,621 which is considerably more than could be counted with the Exponential Autodilution. Furthermore, the known solids loading of this dispersion is 1%.

Since all the particles in the dispersion are countable, the solids volume recovery should be 100%. The 780 recovered 94.4% of the solids volume. Both the particle concentration and % volume recovery numbers confirm the accuracy the Two-Stage Autodilution approach. Overall the Two-Stage Autodilution diluted the sample by a factor of 2,600.

In conclusion, the Two-Stage Autodilution is shown to be accurate. It works best with extremely concentrated dispersions. The system is able to monitor and change dilution factors so has to handle samples that change with time, for example, sample undergoing homogenization. It provides a quick and accurate way to obtain large number of particle counts so as to generate statistically accurate tail information.